Observation | Possible causes | Recommended solutions |
---|---|---|
Droplet containment vacuum not functioning | Worn O-ring in retainer | Replace the O-ring. Contact FMCF staff. |
Outer sleeve is not seated in the retainer | 1. Loosen the retainer 2. Push the outer sleeve up into the retainer until seated. 3. Tighten the retainer | |
Outer sleeve is not on the sample injection tube | Replace the outer sleeve. 1. Loosen the retainer. 2. Slide the outer sleeve over the sample injection tube until it is seated. 3. Tighten the retainer. | |
Waste line is pinched, preventing proper aspiration | Check the waste line. | |
Waste tank is full | Empty the waste tank. |
This is the most commonly asked question. To answer it we need to know the following information:
- What is the approximate percentage of the population(s) you wish to sort?
- How many total cells do you want back?
- Do you want stringent purification or enrichment?
- How fragile or large are your cells?
All of these parameters influence the yield and purity of cell sorting. Here is an example: Let’s say the subset you want sorted is 20% of the total and you need 2 million cells for your experiment. In theory you would need to run 10 million cells through the sorter (10 million x 20% = 2 million). However, the actual yield is usually 75-95% of this theoretical yield, due to abort rates (caused by sort conflicts) and also the quality of your sample.
Therefore, we recommend that you bring 25-50% more cells to the sorter than you would need if the actual yield were 100% (based on the abundance of your target cells).
Observation | Possible causes | Recommended solutions |
---|---|---|
Sample tube not fitting on SIP | Sample tube other than BD Falcon tubes used | Use BD Falcon 12 x 75-mm sample tubes. Tubes must be polystyrene NOT polypropylene |
Worn Bal seal | Replace the Bal seal. Contact FMCF staff |
The answer to this question depends primarily on your goal (enrichment vs stringent purification), the nature of your cells (fragile cells or large cells need to be sorted at lower pressures and speeds with a larger nozzle), and the concentration of your sample (more dilute samples will take longer to sort).
For example, on the MoFlo cell sorter, we use the 100µm nozzle at 30psi. For primary lymphocytes we run samples at maximum flow rates of 10,000-13,000/second. For cell lines, we run samples 400- 4,000/second depending on the size of the cells. With the Aria cell sorters, we can run primary lymphocytes at 6,000-8,000/second with the 100 µm nozzle at 30psi, or at 9,000-10,000/second with the 70 µm nozzle at 60psi.
Again, for cell lines the flow rates will vary a lot depending on the size of cells. For example, if your cells are run at 10,000/second, it would take approximately 30 minutes to run a total of 18 million cells through the sorter. Using the above scenario, your theoretical yield is 18 million x 20% = 3.6 million. If you require stringent purity or your sample is of poor quality (low viability, clumping, etc), your actual yield will be lower.
It is REQUIRED that your cells be filtered through at 70um nylon mesh, preferably right before running on the sorter. We can provide the mesh if necessary.
Observation | Possible causes | Recommended solutions |
---|---|---|
Rapid sample aspiration | Support arm is to the side | Place the support arm under the sample tube |
Droplet containment module is failing | Call your service representative. Contact FMCF staff |
Proper sample preparation is absolutely essential for obtaining good purity and yields. After the final wash, cells should be re-suspended at concentrations of 20-30 million (for primary cells) or 5-10 million (for cell lines) per mL in FACS Tubes (see below). The media can be any buffered isotonic salt solution (such as PBS) but we recommend Hanks Balanced Salt Solution (HBSS) calcium and magnesium free containing 10 mM HEPES, pH 7.2).
Most cells benefit from having some protein in the media, such as 1-2% FBS, calf serum or BSA. Higher serum concentrations can lead to cell aggregation and clogging. For cell line samples, we recommend also adding 1mM EDTA. We recommend using buffers that do not contain phenol red because phenol red causes an increase in background fluorescence, and reduces sensitivity. This effect is particularly bad for dim green signals.
For sticky cells: Increase the concentration of the EDTA to 5mM and use FBS that has been dialyzed against Ca/Mg++ free PBS or 0.5% BSA. Some cell types, however, can be sensitive to high concentrations of EDTA.
For adherent cells: To detach adherent cells, trypsin (or other detachment buffer) is often quenched with media or buffer containing serum. This reintroduces cations that can promote aggregation. Use FBS that has been dialyzed against Ca/Mg++ free PBS or increase the EDTA concentration to 5mM or higher (Some cell types, however, can be sensitive to high concentrations of EDTA.). Accutase and Accumax are cell detachment solutions that can be used to generate single cell suspensions.
NOTE: Accutase can alter some cell surface epitopes; the effect will need to be determined empirically for the epitopes being evaluated.
For samples with a high percentage of dead cells: DNA from dead cells can cause cell clumping. DNAse I in the presence of magnesium chloride will help reduce cellular aggregation.
- Treat cells for 15 to 30 minutes in a solution of 100 µg/mL DNAse and 5 mM MgCl2 in HBSS at room temperature.
- Wash the cells once in HBSS with 5 mM MgCl2.
- Gently resuspend the cells in HBSS with 5 mM MgCl2 and add 25-50 µg/mL DNAse (as a maintenance dose) prior to and during the sort.
Note: DNAse I requires at least 1 mM magnesium to work effectively, although 5mM is optimal. It is important to minimize the presence of dead cells during this procedure, since actin released from dead cells irreversibly inhibits DNAse I.
You will need to bring an unstained control, single colour compensation controls and you MAY need to bring FMO (Fluorescence Minus One) gating controls. You may want to consider an FMO control if you are using more than 5 colours simultaneously and if your populations of interest cannot be discretely separated from background/negatives.
Please refer to the table below to see an example of a typical sort experiment with three colours.
Control type | Sample Name | Marker + fluorochrome #1 | Marker + fluorochrome #2 | Marker + fluorochrome #3 |
---|---|---|---|---|
Unstained Control | Unstained | - | - | - |
Single colour compensation controls | CD3 - FITC | CD3 - FITC | - | - |
CD4 - PE | - | CD4 - PE | - | |
CD8 - APC | - | CD8 - APC | ||
FMO Control | FMO FITC | - | CD4 - PE | CD8 - APC |
Fluorophores excited by the Yellow/Green laser
If you are using markers such as mCherry, RFP, Tomato Red, Texas-Red, etc, Please contact the Flow Facility to inquire about the specific sorter to book as only a limited number of sorters are equipped with an appropriate laser to detect these signals.
Viability Dyes
We recommend that you add a viability dye such as PI, 7AAD, or LIVE/DEAD fixable dyes (Invitrogen) to your sample to enable exclusion of dead cells during sorting. If you choose to do so, please bring an appropriate single color compensation control. We suggest the following: Heat killed cells (65oC for 15 minutes and then mix with live cells) followed by staining with viability dye.
BD Falcon makes FACS Tubes of various kinds. For all cell sorters, order 5 ml sterile polypropylene tubes (catalogue number BD2063). For non-sterile and analytical samples order catalogue number BD2002 (polypropylene) or BD2052 (polystyrene).
For abundant populations we recommend 50 mL (MoFlo) or 15 mL (FACS Aria) Falcon Tubes, or 5 mL Falcon tubes. We can also sort into a range of microtiter plates, including from 96-well to 8-well, microscope slides or Petri dishes. Please talk to us about your requirements and we will advise you on the best choice for your experiment.
Observation | Possible causes | Recommended solutions |
---|---|---|
No events in acquisition display and RUN button is green | Threshold is not set to the correct parameter (usually FSC) | Set the threshold to the correct parameter for your application |
Threshold level is too high | Lower the threshold level | |
PMT voltage for threshold parameter is set too low | Set the PMT voltage higher for the threshold parameter | |
Gating issue | Ensure gating strategy is appropriate | |
Air in the sheath filter | Purge the filter | |
Sheath and air line not connected to the sheath tank | Ensure sheath and air line are connected to the sheath tank | |
O-ring is not in position in the sheath tank | Reposition O-ring in sheath tank | |
Cracked tube/ Bal seal (tube to SIP connection) problem | Replace tube | |
No sample in the tube | Add sample to the tube or install a new sample tube | |
Sample is not mixed properly | Mix the sample to suspend cells | |
Waste tank is full | Empty the waste tank | |
PMT voltages set too low or too high for display parameter | Reset the PMT voltages | |
Too few events are displayed | Increase the number of events to display | |
Sample injection tube is clogged | Remove the sample tube to allow backflushing and prime the instrument several times. Clean instrument. Contact FMCF staff if issue is not resolved | |
Laser is not warmed up | Wait the recommended amount of time for the laser to warm up. - 30 min for the 488-nm (blue) - 30 min for the 355-nm (UV) - 15 min for the 405-nm (violet) - 20 min for the 633-nm (red) | |
Laser delay is set incorrectly | Contact FMCF staff | |
Laser is not functioning | Verify the malfunction by changing the threshold to an alternative laser while running the appropriate sample. If unsuccessful, contact FMCF staff |
We recommend 100% serum or HEPES buffered HBSS.
The sorter uses PBS as its sheath fluid. Mixing large quantities of PBS with buffer containing calcium chloride may produce a precipitate if calcium phosphate on the cell membranes. This can adversely affect viability. For this reason, it is preferable to sort large numbers of cells into 100% serum, and not into regular culture medium. HEPES buffered HBSS without calcium is an alternative. If you must sort into medium, remember that the pH will rise with time, and it’s best to start with it on the acid side of neutral.
Observation | Possible causes | Recommended solutions |
---|---|---|
No events in acquisition display and RUN button is orange | RUN is not activated | Press the RUN button |
Sample tube is not installed or is not properly seated | Install the sample tube correctly on the SIP | |
Sample tube is cracked | Replace the sample tube | |
Sheath container is not pressurized | 1. Ensure that the sheath container lid and all connectors are securely seated. 2. Inspect the O-ring and replace it if necessary. Contact FMCF staff. | |
Bal seal is worn | Replace the Bal seal. Contact FMCF staff | |
Air leak at sheath container | Ensure that the sheath container lid and all connectors are securely seated | |
Sheath container is empty | Fill the sheath container | |
Air in sheath filter | Purge the filter |
Observation | Possible causes | Recommended solutions |
---|---|---|
No fluorescent signal | Incorrect fluorochrome assignment | Make sure the cytometer configuration in the software matches the optical filters in the cytometer |
Wrong filter is installed | Make sure the appropriate filter is installed for each fluorochrome | |
Laser is not functioning | Verify the laser malfunction by changing the threshold to an alternative laser while running the appropriate sample. If unsuccessful, contact FMCF staff |
Observation | Possible causes | Recommended solutions |
---|---|---|
High event rate | Air bubble in the sheath filter or flow cell | Remove the air bubble from filter |
Sample is too concentrated | Dilute the sample | |
Sample flow rate is set on HI | Set the sample flow rate to MED or LO | |
Threshold level is too low | Increase the threshold level | |
PMT voltage for the threshold parameter set too high | Set the PMT voltage lower for the threshold parameter |
Observation | Possible causes | Recommended solutions |
---|---|---|
Low event rate | Threshold level is too high | Lower the threshold level |
PMT voltage for the threshold parameter is set too low | Set the PMT voltage higher for the threshold parameter | |
Sample is not adequately mixed | Mix the sample to suspend the cells | |
Sample is too diluted | Concentrate the sample. If the flow rate setting is not critical to the application, set the flow rate switch to MED or HI | |
Sample injection tube is clogged | Remove the sample tube to allow backflushing. If the event rate is still erratic, clean the sample injection tube |
Observation | Possible causes | Recommended solutions |
---|---|---|
Erratic event rate | Sample tube is cracked | Replace the sample tube |
Bal seal is worn | Replace the Bal seal. Contact FMCF staff | |
Sample injection tube is clogged | Remove the sample tube to allow backflushing and prime the instrument several times. Contact FMCF staff if the issue is not resolved | |
Contaminated sample | Prepare the specimen again. Ensure that the tube is clean | |
Sheath filter is dirty | Replace the filter. Contact FMCF staff |
Observation | Possible causes | Recommended solutions |
---|---|---|
Distorted scatter parameters | Cytometer settings are improperly adjusted | Optimize the scatter parameters |
Air bubble in sheath filter or flow cell | Purge the air from the filter | |
Flow cell is dirty | Perform the system flush procedure. Contact FMCF staff | |
Air leak at sheath container | Ensure that the sheath container lid is tight and all connectors are secure | |
Hypertonic buffers or fixative | Replace the buffers and fixative |
Observation | Possible causes | Recommended solutions |
---|---|---|
Excessive amount of debris in display | Threshold level is too low | Increase the threshold level |
Sheath filter is dirty | Replace the filter. Contact FMCF staff | |
Flow cell is dirty | Prime the instrument several times. Contact FMCF staff | |
Dead cells or debris in sample | Examine the sample under a microscope | |
Sample is contaminated | Re-stain the sample, ensure tube is clean | |
Stock sheath fluid is contaminated | Contact FMCF staff |
Observation | Possible causes | Recommended solutions |
---|---|---|
High CV | Air bubble in sheath filter or flow cell | Purge the filter |
Sample flow rate is set too high | Set the sample flow rate lower | |
Air leak at sheath container | Ensure that the sheath container lid is tight and all connectors are secure | |
Flow cell is dirty | Prime the instrument several times. Contact FMCF staff | |
Poor sample preparation | Repeat sample preparation | |
Sample not diluted in same fluid as sheath fluid | Dilute the sample in the same fluid as you are using for sheath |
Observation | Possible causes | Recommended solutions |
---|---|---|
Poor QC results | Air bubble or debris in flow cell | Prime the fluidics system |
Old or contaminated QC particles | Make new QC samples and perform the quality control procedure again | |
Sample not diluted in same fluid as sheath fluid | Dilute the sample in the same fluid as you are using for sheath | |
Laser not warmed up | Wait the recommended amount of time for the laser to warm up. - 30 min for the 488-nm (blue) - 30 min for the 355-nm (UV) - 15 min for the 405-nm (violet) - 20 min for the 633-nm (red) | |
Laser not functioning | Contact FMCF staff | |
Optical alignment problem | Contact FMCF staff |