ObservationPossible causesRecommended solutions
No events in acquisition display and RUN button is greenThreshold is not set to the correct parameter (usually FSC)Set the threshold to the correct parameter for your application
Threshold level is too highLower the threshold level
PMT voltage for threshold parameter is set too lowSet the PMT voltage higher for the threshold parameter
Gating issueEnsure gating strategy is appropriate
Air in the sheath filterPurge the filter
Sheath and air line not connected to the sheath tankEnsure sheath and air line are connected to the sheath tank
O-ring is not in position in the sheath tankReposition O-ring in sheath tank
Cracked tube/ Bal seal (tube to SIP connection) problemReplace tube
No sample in the tubeAdd sample to the tube or install a new sample tube
Sample is not mixed properlyMix the sample to suspend cells
Waste tank is fullEmpty the waste tank
PMT voltages set too low or too high for display parameterReset the PMT voltages
Too few events are displayedIncrease the number of events to display
Sample injection tube is cloggedRemove the sample tube to allow backflushing and prime the instrument several times. Clean instrument. Contact FMCF staff if issue is not resolved
Laser is not warmed upWait the recommended amount of time for the laser to warm up.
- 30 min for the 488-nm (blue)
- 30 min for the 355-nm (UV)
- 15 min for the 405-nm (violet)
- 20 min for the 633-nm (red)
Laser delay is set incorrectlyContact FMCF staff
Laser is not functioningVerify the malfunction by changing the threshold to an alternative laser while running the appropriate sample. If unsuccessful, contact FMCF staff