Observation | Possible causes | Recommended solutions |
---|---|---|
Erratic event rate | Sample tube is cracked | Replace the sample tube |
Bal seal is worn | Replace the Bal seal. Contact FMCF staff | |
Sample injection tube is clogged | Remove the sample tube to allow backflushing and prime the instrument several times. Contact FMCF staff if the issue is not resolved | |
Contaminated sample | Prepare the specimen again. Ensure that the tube is clean | |
Sheath filter is dirty | Replace the filter. Contact FMCF staff |