ObservationPossible causesRecommended solutions
No events in acquisition display and RUN button is orangeRUN is not activatedPress the RUN button
Sample tube is not installed or is not properly seatedInstall the sample tube correctly on the SIP
Sample tube is crackedReplace the sample tube
Sheath container is not pressurized1. Ensure that the sheath container lid and all connectors are securely seated.
2. Inspect the O-ring and replace it if necessary. Contact FMCF staff.
Bal seal is wornReplace the Bal seal. Contact FMCF staff
Air leak at sheath containerEnsure that the sheath container lid and all connectors are securely seated
Sheath container is emptyFill the sheath container
Air in sheath filterPurge the filter