Observation | Possible causes | Recommended solutions |
---|---|---|
No events in acquisition display and RUN button is orange | RUN is not activated | Press the RUN button |
Sample tube is not installed or is not properly seated | Install the sample tube correctly on the SIP | |
Sample tube is cracked | Replace the sample tube | |
Sheath container is not pressurized | 1. Ensure that the sheath container lid and all connectors are securely seated. 2. Inspect the O-ring and replace it if necessary. Contact FMCF staff. | |
Bal seal is worn | Replace the Bal seal. Contact FMCF staff | |
Air leak at sheath container | Ensure that the sheath container lid and all connectors are securely seated | |
Sheath container is empty | Fill the sheath container | |
Air in sheath filter | Purge the filter |